Simultaneous Quantitation of Two Potential Genotoxic Impurities In Imatinib by Liquid Chromatography Mass Spectrometry

Kotapati Nalini et al.
2/1/2017

A new, simple, accurate and sensitive method was developed for the quantification of two potential genotoxic impurities 1-(2-methyl-5-nitro-phenyl)guanidine nitrate and  N-(2-methyl-5-nitro-phenyl)-4-(3-pyridyl)pyrimidin-2-amine at low level (2 ppm) in Imatinib using liquid chromatographic mass spectrometry. The chromatographic separation was achieved on Kromasil 100-3.5-C18 (150 x 2.1) mm column with gradient programme and elution was monitored by mass spectrometer in selective ion monitoring mode using electrospray ionization. The LOD and LOQ values found to be 0.2 ppm and 0.6 ppm for both the impurities with respect to the test concentration 5 mg/ml. The method was linear (r2>0.99), precise (RSD